B-5 Resistance
200Ω
2KΩ
20KΩ
200KΩ
2MΩ
20MΩ
200MΩ
0.01Ω
0.1Ω
1Ω
10Ω
100Ω
1KΩ
10KΩ
Range Resolution Accuracy
(0.3% +1)
(0.5% +10)
(0.5% +1)
[5.0%(-1000) +10]
(0.3% +3)
Overload protection: All ranges are 250VDC
or AC RMS.
Caution:
1) At 200MΩ range, test lead is short circuit, the
LCD display 1000 digits, deduct the 1000
digits from the measured reading during
measuring.
2) At 200Ω range, short the test lead first to display
the resistance value of the test leads. Deduct
that resistance value from the measured reading
to obtain the being measured object correct
value .
B-6 Capacitance
Testing Signals: About 400Hz, 40mVrms.
2nF
20nF
200nF
2µF
20µF
0.1pF
1pF
10pF
0.1nF
1nf
Range Resolution Accuracy
(4% +20)