Introduction
Unpacking Inspection
Safety Information
Digital Multimeter UT56 (hereafter referred to as "the
Meter") is 20000-count hand-held instruments
featuring stable performance, versatile functions, high
degree of reliability&accuracy. It is designed with
large-scale integrated circuits, dual integral A/D
converter as well as overloaded protection. The series
can measure AC/DC voltage, AC/DC current,
resistance, capacitance, frequency, transistor, diode
and continuity, making the series a perfect tool for
users.
Your Meter's Features
" "
" "
Direct Current Current (DC Current)
2mA
20mA
200mA
20A
0.1
1
10
1A
Range Resolution
(0.8% +5)
Accuracy
(2% +10)
(0.5% +5)
Overload protection: Below 200mA with 0.3
A/250V Fused. No fuse protection on 20A.
Measuring voltage drop: 200mV for full ranges.
Max current input: 20A (>10A: measurement
time <15seconds )
Alternate Current Current (AC Current)
2mA
20mA
200mA
20A
0.1
1
10
1A
Range Resolution
(1.2% +10)
Accuracy
(2.5% +10)
(0.8% +10)
Overload protection: Below 200mA with0.3A/
250V Fused. No fuse protection on 20A.
Max current input:20A (>10A: measurement
time <15seconds )
Measuring voltage drop: 200mV for full ranges.
Display: Average value(RMS of Sine Wave).
A
A
A
A
A
A
Resistance
200Ω
2KΩ
20KΩ
200KΩ
2MΩ
20MΩ
200MΩ
0.01Ω
0.1Ω
1Ω
10Ω
100Ω
1KΩ
10KΩ
Range Resolution Accuracy
(0.3% +1)
(0.5% +10)
(0.5% +1)
[5.0%(-1000) +10]
(0.3% +3)
Overload protection: 250VDC or AC RMS for
Caution:
1) At 200M
Ω range, test lead is short circuit, the
LCD display 1000 digits, subtract the 1000
digits from the measured reading during
measuring.
2) At 200
Ω range, short the test lead first to display
the resistance value of the test leads. subtract
that resistance value from the measured reading
to obtain the being measured object's correct
value .
Capacitance
Testing Signals: About 400Hz, 40mVrms.
2nF
20nF
200nF
2μF
20μF
0.1pF
1pF
10pF
0.1nF
1nF
Range Resolution Accuracy
(4% +20)
Frequency
Range Resolution Accuracy
20kH
z
1H
z
(1.5% +5)
Input sensitivity: <200mVrms, Max. input
30Vrms.
Overload protection: 250Vrms.
Diode and Continuity Beeper
Range
Remark
Measuring Condition
Display Diode
Forward Voltage
Value, unit is "V"
Forward DC Current abt
1mA, Backward DC
Voltage about 3.0V
Beeper sounds when
Resistance < 50Ω.
Display the nearest
value, unit is "kΩ"
Voltage at open Circuit
about 3.0V
Overload protection: 250V DC or AC RMS.
Unpack the carton and take out the Meter. Check
the following accessories for any missing or
damaged component. Please contact your
dealership immediately if any missing or damaged
component is found.
Operating Manual------1pc
Test Leads --------------1 pair
Holster(Optional)
UT56 is designed and manufactured in
compliance with: IEC1010-1, CATⅠ1000V, CATⅡ
600V and CATⅢ300V, Double Insulation and
Pollution Degree.
Warnings
Use the meter as specified in the manual,
otherwise the protection offered by the Meter may
be impaired.
Ɣ'RQRWXVHWKH0HWHUZLWKEDFNFRYHURSHQHGLW
may cause electric shock.
Ɣ6HWWRDSURSHUUDQJHIRU\RXUPHDVXUHPHQW
Ɣ&KHFNWKHWHVWOHDGVIRUDQ\GDPDJHGLQVXODWLRQ
or broken wires.
Ɣ,QVHUWUHGDQGEODFNWHVWOHDGVLQWRSURSHULQSXW
terminals and ensure good contact between them.
Ɣ'RQRWLQSXWVLJQDOEH\RQGWKHUDWHGYDOXHVRI
the Meter, otherwise it may cause electric shock
or damage to the Meter.
Ɣ3OHDVHXVHUHSODFHPHQWSDUWVZLWKWKHVDPH
model or similar electrical specifications.
Ɣ7RDYRLGHOHFWULFVKRFN'RQRWDSSO\DQ\YROWDJH
above 1000V between COM and grounding.
Ɣ3OHDVHWDNHFDXWLRQZKHQZRUNLQJYROWDJHJR
above 60VDC or 30VAC RMS.
Ɣ7RHQVXUHWKHDFFXUDF\SOHDVHUHSODFHWKH
batteries as soon as
" " appears.
Ɣ6KXWRIIWKHSRZHUWLPHO\DIWHUWKHPHDVXUHPHQWV
complete. Take out the batteries if not used for a
long time.
Ɣ'RQRWXVHWKH0HWHULQSODFHVH[SRVHGWRKLJK
temperature, high moisture. The performance of
the Meter may be compromised if
moisture-affected.
Ɣ7RSUHYHQWGDPDJHWRWKH0HWHURUSHUVRQDO
injury, do not alter internal wiring randomly.
Caution
0ȍIRUDOOUDQJHV
0ȍIRUDOOUDQJHV
250VDC or AC RMS for
250VDC or AC RMS for
200mV range. 750VRMS or 1000Vp-p for other
ranges.
200mV range. 750VRMS or 1000Vp-p for other
ranges.
all ranges.