UNI-T UT209 Spezifikationen

Stöbern Sie online oder laden Sie Spezifikationen nach Multimeter UNI-T UT209 herunter. UNI-T UT209 Specifications Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - OPERATING MANUAL

Model UT209OPERATING MANUAL

Seite 2

International Electrical SymbolsAC (Alternating Current)DC (Direct Current)AC or DCGroundingDouble InsulatedWarning. Refer to the Operating ManualDefi

Seite 3

Figure 1The Meter Structure (see figure 1)Hand Guards: to protect user’s handfrom touching the dangerous area.Trigger: press the lever to open thetra

Seite 4 - Overview

Press SELECT button to select the alternate functions includingV , andButton Operation PerformedSELECTPMAX/PMINPress HOLD to enter the Ho

Seite 5 - Unpacking Inspection

Button Operation PerformedHOLDHzZEROPress ZERO to display the present value as stored value anddisplay zero. After that, all the measureme

Seite 6 - Safety Information

SELECT MAX/MINHOLDHzZERO%Hz44A1000A40A1000AN/AN/AN/AN/AN/AN/AN/ARotary SwitchPositionsVFunctional ButtonsN/AN/AThe Effectiveness of Functional Button

Seite 7 - Rules For Safe Operation

Figure 2Display Symbols (see figure 2)Model UT209: OPERATING MANUAL14

Seite 8

Number Symbol Meaning1234567891011Hz,kHz,MHz,k ,MPMAXPMINZEROHz: Hertz. The unit of frequency.KHz: Kilohertz. 1 x 103 or 1000 her

Seite 9

Number Symbol Meaning1213141516171819The battery is low. Warning: To avoid false readings, whichcould lead to possible electr

Seite 10 - Model UT209: OPERATING MANUAL

A.DC/AC Voltage Measurement (see figure 3)To avoid harms to you or damages to the Meter from eletric shock, do notattempt to measure voltages higher t

Seite 11

To measure DC/AC voltages, connect the Meter as follows:Insert the red test lead into the V Hz terminal and black test lead into the COMterminal.Set

Seite 12 - Functional Buttons

Table of ContentsTitle PageOverviewUnpacking InspectionSafety InformationRules For Safe OperationInternational Electrical SymbolsThe Meter StructureFu

Seite 13

B.Measuring Resistance (see figure 4)To avoid damages to the Meter or to the devicesunder test, disconnect circuit power anddischarge all the high-vol

Seite 14

3. Connect the test leads across with the object being measured.The measured value shows on the display.NoteTo obtain a more precise reading, you coul

Seite 15 - (see figure 2)

C. Testing Diodes (see figure 5)To avoid damages to the Meter or to the devicesunder test, disconnect circuit power anddischarge all the high-voltage

Seite 16

Insert the red test lead into the V Hz terminal and black test lead into the COMterminal.Set the rotary switch to Press SELECT to switch to measure

Seite 17

D. Testing for Continuity (see figure 6)To avoid damages to the Meter or to the devicesunder test, disconnect circuit power anddischarge all the high-

Seite 18 - Measurement Operation

a circuit under test is higher than 100NoteWhen continuity testing has been completed, disconnect the connection betweenthe testing leads and the cir

Seite 19

E. Frequency Measurement (see figure 7)To avoid harms to you or damages to the Meterfrom eletric shock, do not attempt to measurevoltages higher than

Seite 20

NoteThe maximum amplitude must not exceed 30Vrms.When frequency measurement has been completed, disconnect the connectionbetween the testing leads and

Seite 21

F. DC Current Measurement (see figure 8)The operating temperature must be 0 ~40when measuring current.The measurement ranges of current are: 40A and

Seite 22

not display zero. When the testing current is greater than 400A , it is notnecessary to zero.Hold the Meter tight, don’t release. The Hall componen

Seite 23

Title Page3032333334353536373838394040414142 G. AC Current MeasurementSleep ModeSpecifications A. General Specifications B. Environmental

Seite 24

Note Press ZERO to display the present value as stored value and display zero. After that, all the measurement result displayed will automatic

Seite 25

G. AC Current Measurement (see figure 9)The operating temperature must be 0 ~40when measuring current.The measurement ranges of current are: 40A a

Seite 26

NoteWhen measuring AC current range, It may have unstable or wrong sensed readings,it will not affect measurement result.Change to AC:Couple AC and Tr

Seite 27

Sleep ModeTo preserve battery life, the Meter automatically turns off if you do not turn the rotaryswitch or press any button for around 30 minutes.Th

Seite 28

l Maximum Voltage between any Terminals and grounding: Refer to different range input protection voltage.l Display: 3 3/4 digits LCD display, Ma

Seite 29

l Battery Life: typically 150 hours (alkaline battery)l Dimensions: 285.3mm x 105mm x 44.5mml Weight: Approximate 533g (battery included)B. Environme

Seite 30

Accurate SpecificationsAccuracy: (a% reading + b digits), guarantee for 1 year.Operating temperature: 23 5Relative humidity: 85%R.HTemperature co

Seite 31

B. AC VoltageRange400mV4V40V400V600VResolution0.1mV0.001V0.01V0.1V1VAccuracy(1.2%+20)(1.2%+5)(1.5%+5)Overload protection600V DC/ACRemarks:Input Impeda

Seite 32

Peak factor: 1.4~2.0, add 1.0% on the stated accuracyPeak factor: 2.0~2.5, add 2.5% on the stated accuracyPeak factor: 2.5~3.0, add 4.0% on the stated

Seite 33 - Sleep Mode

D. Diode TestRange Resolution1mVAccuracy0.5V~0.8V (Opencircuit voltageapprox. 3.0V)Overload protection250VACE. Continuity TestRange Resolution0.1Accur

Seite 34 - Specifications

OverviewThis Operating Manual covers information on safety and cautions. Please read therelevant information carefully and observe all the Warnings an

Seite 35

Remarks:l Input Sensitivity as follows: When 100kHz: 300mV rms When>100kHz: 600mV rmsl Input amplitude a: 300mV a 30V rmsF. Fr

Seite 36 - Accurate Specifications

G. DC CurrentRange40A1000AResolution0.01A1AAccuracy(2%+40)(2%+8)Overload protection 250VACH. AC CurrentRange40A1000AResolution0.01A 1AAccuracy(

Seite 37

MAINTENANCEThis section provides basic maintenance information including battery replacementinstruction.Do not attempt to repair or service your Meter

Seite 38

inflammable and strong magnetic field.To avoid false readings, which couldlead to possible electric shock orpersonal injury, replace the battery asso

Seite 39

To replace the battery:.1. Turn the Meter off and remove all the connections from the input terminals2. Turn the Meter’s front case down.3. Remove the

Seite 40

* END *This operating manual is subject to change without notice.Model UT209: OPERATING MANUAL44

Seite 41

Copyright 2007 Uni-Trend Group Limited.All rights reserved.Manufacturer:Uni-Trend Technology (Dongguan) LimitedDong Fang Da DaoBei Shan Dong Fang Indu

Seite 42 - MAINTENANCE

Item Description Qty1234English Operating ManualTest LeadTool Box9

Seite 43

Safety InformationThis Meter complies with the standards IEC61010: in pollution degree 2, overvoltagecategory (CAT. II 600V, CAT. III 300V) and double

Seite 44

Rules For Safe OperationTo avoid possible electric shock or personal injury, and to avoid possibledamage to the Meter or to the equipment under test,

Seite 45

changeover of range shall be made during measurement is conducted toprevent damage of the Meter.Do not carry out the measurement when the Meter’s back

Seite 46

When servicing the Meter, use only the same model number or identicalelectrical specifications replacement parts.The internal circuit of the Meter sha